How often should I calibrate the i500 scanner?

The calibration process is essential in producing precise 3D models. It is recommended to perform the calibration process periodically.

– Calibration is required when:

The quality of the 3D model is not reliable or accurate as compared to previous results.
Environmental conditions such as temperature have changed.
Calibration period has expired.
You can set the calibration period as detailed in Menu > Settings > Calibration Period(Days). The default calibration period is 14 days.

– Precautions

The calibration panel is a delicate component. Do not touch the panel directly. When the calibration process is not performing properly, check the panel. If the calibration panel is contaminated, please contact Protec Dental Laboratories.

– How to calibrate the i500

  • Turn on the i500 and connect the device to the software.
  • Run Calibration Wizard from Menu > Settings > Calibration.
  • Prepare the calibration tool and the i500.
  • If the scanner tip is attached, please remove it from the handpiece.

How often should I calibrate the i500 scanner? 1

  • Turn the dial of the calibration tool to the position (1).

How often should I calibrate the i500 scanner? 2

  • Remove the cover from the calibration tool and put the i500 handpiece into the calibration tool.

How often should I calibrate the i500 scanner? 3

  • Click Next to start the calibration process.

How often should I calibrate the i500 scanner? 4

  • When the handpiece is mounted in the correct position, the system will automatically acquire the data at position(1).

How often should I calibrate the i500 scanner? 5

  • Repeat the same for position (2) to position (8) and finally, (LAST)

How often should I calibrate the i500 scanner? 6

  • When data acquisition is completed at the (LAST) position, the system will automatically calculate and show the calibration results.

How often should I calibrate the i500 scanner? 7